FDA Adverse Event Malfunction Summary report: N

ACCU-CHEK ® SPIRIT COMBO

MDR report key: 3103111 · Received May 9, 2013

Report

Report Number
2183996-2013-00818
Event Type
Malfunction
Date Received
May 9, 2013
Date of Event
April 28, 2013
Report Date
June 12, 2013
Manufacturer
ROCHE DIABETES CARE AG
Product Code
LZG
PMA / PMN Number
NA
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

THE COMPLAINT CANNOT BE VERIFIED. THE PRODUCT MEETS THE SPECIFICATION REGARDING THE CUSTOMER'S ALLEGATION. THE DELIVERY ACCURACY OF THE INSULIN PUMP WAS TESTED WITHIN THE PUMP DRIVE TEST ON THE DIAGNOSTIC TEST SYSTEM AND MEETS THE SPECIFICATIONS. THE TECHNICAL INVESTIGATION GAVE NO EVIDENCE THAT THE DEVICE DID CAUSE OR CONTRIBUTE TO THE CONDITION REPORTED BY THE PATIENT. THE ALARM FUNCTIONS OF THE PUMP WERE TESTED ON THE DIAGNOSTIC TEST SYSTEM AND MEET THE SPECIFICATIONS.

Additional Manufacturer Narrative · 1

THIS INCIDENT OCCURRED OUTSIDE THE UNITED STATES. INFORMATION CONTAINED WITHIN THIS REPORT IS ALL THAT IS AVAILABLE AT THIS TIME. IF FURTHER INFORMATION IS OBTAINED IT WILL BE PROVIDED IN THE SUPPLEMENTAL REPORT.

Description of Event or Problem · 1

PATIENT REPORTED HE HAS EXPERIENCED HYPERGLYCEMIA OF UP TO 400 MG/DL SINCE (B)(6) 2013, AND THE INFUSION DEVICE HAS NOT DISPLAYED ANY ALERT/ERROR MESSAGES. HE CHANGED THE INFUSION SET AND DELIVERED INSULIN VIA THE INFUSION DEVICE, BUT THIS DID NOT RESOLVE THE ISSUE. FOLLOWING HIS PHYSICIAN'S RECOMMENDATION, HE DELIVERED INSULIN VIA PEN AND WAS ABLE TO LOWER HIS BLOOD GLUCOSE TO HIS NORMAL RANGE (APPROXIMATELY 120 MG/DL). THE INFUSION DEVICE WAS REQUESTED FOR EVALUATION. HE DID NOT REQUIRE TREATMENT FROM A HEALTHCARE PROFESSIONAL OR SECOND PARTY TO ADDRESS THE ISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
203014 ACCU-CHEK ® SPIRIT COMBO INSULIN INFUSION PUMP LZG ROCHE DIABETES CARE AG 00700006863 NA

Patients

Seq Age Sex Outcome Treatment
1 023 YR