FDA Adverse Event Malfunction Summary report: N

MEGA SUTURECUT NEEDLE DRIVERINSTRUMENT

MDR report key: 3103081 · Received May 9, 2013

Report

Report Number
2955842-2013-01570
Event Type
Malfunction
Date Received
May 9, 2013
Date of Event
April 8, 2013
Report Date
April 9, 2013
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VT
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. ENGINEERING CONFIRMED THE REPORTED COMPLAINT. ONE GRIP CLOSE CABLE WAS BROKEN AT THE DISTAL IDLERS PULLEY. THE IDLER PULLEY SPUN FREELY BUT HAD DAMAGE ON THE SURFACE AND EDGE. ENGINEERING CONCLUDED THE DAMAGE WAS LIKELY DUE TO MISHANDLING. THE CABLE SEGMENT STUCK OUT AT THE INSTRUMENT'S WRIST. OTHER CABLES AT THE WRIST WERE NOT DAMAGED. THE ENDOWRIST INSTRUMENTS INSTRUCTIONS FOR USE (IFU) SPECIFICALLY STATES: GENERAL PRECAUTIONS AND WARNINGS HANDLE INSTRUMENTS WITH CARE. AVOID MECHANICAL SHOCK OR STRESS THAT CAN CAUSE DAMAGE TO THE INSTRUMENTS. DO NOT USE AN INSTRUMENT TO CLEAN DEBRIS FROM ANOTHER INSTRUMENT INTRAOPERATIVELY. THIS MAY RESULT IN DAMAGE TO THE INSTRUMENTS OR OTHER UNINTENDED CONSEQUENCES, SUCH AS DISCONNECTION OF THE INSTRUMENT TIP. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER; THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING REPROCESSING THE MEGA SUTURECUT NEEDLE DRIVER INSTRUMENT THAT THE WIRE BROKE. THERE WERE NO MISSING PIECES OR FALLEN PIECES REPORTED. NO PATIENT HARM, ADVERSE OUTCOME, OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
204122 MEGA SUTURECUT NEEDLE DRIVERINSTRUMENT ENDOSCOPIC INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420309-01 M10121030 605

Patients

Seq Age Sex Outcome Treatment
1 DA VINCI S SYSTEM INSTRUMENTS AND ACCESSORIES