ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2013-06110
- Event Type
- Injury
- Date Received
- May 9, 2013
- Date of Event
- April 14, 2013
- Report Date
- April 14, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- PATIENT
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
ON (B)(6) 2013 THE REPORTER CONTACTED ANIMAS ALLEGING THAT THE PATIENT'S BLOOD GLUCOSE (BG) WAS 558 MG/DL AND THE PUMP ALARMED THAT IT HAD EXCEEDED THE DAILY MAXIMUM LIMIT (MAX TDD). CUSTOMER TECHNICAL SUPPORT (CTS) REVIEWED THE PUMP HISTORY WITH THE REPORTER AND CONFIRMED ALL THE EVENING BOLUSES PROGRAMMED HAD BEEN RECEIVED. THE REPORTER NOTED THAT THE PATIENT HAD BEEN BOLUSING MORE DUE TO HIGH CARBOHYDRATE MEALS AND CONFIRMED THAT THE PATIENT'S BG RESPONDED TO BOLUSES. CTS ADVISED THE REPORTER THAT AT MIDNIGHT THE TDD WOULD RE-SET AND BOLUSES COULD BE DELIVERED AGAIN. CTS ALSO REVIEWED CHANGING THE MAX TDD LIMITS ON THE PUMP. CTS ADVISED THE REPORTER IF BGS DO NOT COME DOWN OR IF THERE ARE ANY OTHER ISSUES, TO CALL BACK. THE PUMP WAS FOUND TO BE ALARMING APPROPRIATELY ACCORDING TO SETTINGS AND THERE WAS NO DEFECT FOUND DURING TROUBLESHOOTING. THIS COMPLAINT IS BEING REPORTED BECAUSE THE PATIENT ALLEGEDLY EXPERIENCED HYPERGLYCEMIA WHILE USING INSULIN PUMP THERAPY DUE TO EXCEEDING THE MAXIMUM DAILY INSULIN DELIVERY LIMITS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 203432 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 14 YR | Life Threatening |