HOMECHOICE PRO
Report
- Report Number
- 1416980-2013-11837
- Event Type
- Malfunction
- Date Received
- May 9, 2013
- Date of Event
- March 24, 2013
- Report Date
- April 19, 2013
- Manufacturer
- BAXTER HEALTHCARE - LARGO
- Product Code
- FKX
- PMA / PMN Number
- K102936
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TW
- Reporter Occupation
- PATIENT
Narratives
(B)(4). THE DEVICE WAS RETURNED TO BAXTER, AND THE EVALUATION IS COMPLETE. THIS COMPLAINT IS AN ANCILLARY SERVICE EVENT. THE REPORTED DIFFICULTY OF AN IIPV EVENT WAS CONFIRMED THROUGH AN EVENT HISTORY LOG REVIEW. THE CAUSE WAS DETERMINED TO BE FALSE EMPTY DETECT AND USE ERROR, CLINICIAN INAPPROPRIATELY SET MINIMUM DRAIN VOLUME PERCENT SETTING TOO LOW. HOMECHOICE APD SYSTEMS TRAINER'S GUIDE. SECTION 12 "PROGRAMMING A PRESCRIPTION" IN 12.3.3 ON PAGE 12-6 GIVES THE WARNING THAT "IF YOU SET THE MINIMUM DRAIN VOLUME% TOO LOW, AN INCOMPLETE DRAIN COULD RESULT FOR THE PATIENT. THIS COULD CAUSE AN INCREASED INTRAPERITONEAL VOLUME (IIPV) SITUATION". IF ANY ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW-UP WILL BE SENT.
DURING EVALUATION OF A RETURNED HOMECHOICE MACHINE, ONE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED WHICH OCCURRED IN THE THERAPY INITIATED ON (B)(6) 2013 AT 20:15:03. DURING NIGHT DRAIN CYCLE TWO, THE PATIENT'S ULTRAFILTRATION READING WAS 896ML, INDICATING THE HOME PATIENT (HP) DRAINED 896ML MORE THAN THEIR MAXIMUM PROGRAMMED FILL VOLUME OF 1200ML. THIS INFORMATION MEETS IIPV CRITERIA. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 203774 | HOMECHOICE PRO | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - LARGO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |