PROMUS ELEMENT ¿
Report
- Report Number
- 2134265-2013-03127
- Event Type
- Death
- Date Received
- May 9, 2013
- Date of Event
- April 12, 2013
- Report Date
- April 13, 2013
- Manufacturer
- BOSTON SCIENTIFIC - GALWAY
- Product Code
- NIQ
- PMA / PMN Number
- SIMILAR
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
Narratives
AGE AT TIME OF EVENT: (B)(6) OR OLDER. THIS IS A COMBINATION PRODUCT. DEVICE EVALUATED BY MFR: THE COMPLAINT DEVICE WAS NOT RETURNED FOR ANALYSIS AS IT WAS DISPOSED. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE ROOT CAUSE IS ANTICIPATED PROCEDURAL COMPLICATIONS AS THIS EVENT IS A KNOWN PHYSIOLOGICAL EFFECT OF THE PROCEDURE AND IS NOTED WITHIN THE DFU. (B)(4).
IT WAS REPORTED THAT DURING A CORONARY STENTING TREATMENT PROCEDURE, A STENT DAMAGE AND VESSEL DISSECTION OCCURRED. THE 75% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS MID LEFT ANTERIOR DESCENDING (LAD) ARTERY. A 2.50MM X 20MM PROMUS ELEMENT WAS SELECTED FOR TREATMENT. DURING THE PROCEDURE, THE DEVICE WAS UNABLE TO CROSS THE LESION EVEN AFTER SEVERAL ATTEMPTS. AN ANTIDROMIC DISSECTION OCCURRED FROM THE LAD PROXIMAL TO THE LEFT MAIN CORONARY ARTERY (LMCA). THE LAD THEN HAD NO FLOW THUS THE PHYSICIAN REMOVED THE PROMUS ELEMENT SDS AND FOUND THAT THE STENT PROXIMAL EDGE HAD LIFTED. NON-BSC STENTS WERE IMPLANTED IN THE LMCA-LAD PROXIMAL TO TREAT THE DISSECTION. AFTER WHICH, THE LAD AND THE LEFT CIRCUMFLEX BECAME A NO FLOW. PERCUTANEOUS CARDIOPULMONARY SUPPORT (PCPS) SYSTEM WAS PLACED. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THE PATIENT CONDITION IMPROVED. THEN PATIENT DIED ONE DAY AFTER THE PROCEDURE DUE TO BLEEDING FROM PCPS. THE PHYSICIAN FELT THE STENT WAS DAMAGED DUE TO A 90 DEGREE BEND IN THE CALCIFIED LESION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 203213 | PROMUS ELEMENT ¿ | STENT, CORONARY, DRUG-ELUTING | NIQ | BOSTON SCIENTIFIC - GALWAY | H7493911320250 | 14967418 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Death| R |