PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM
Report
- Report Number
- 2024168-2013-02890
- Event Type
- Injury
- Date Received
- May 9, 2013
- Date of Event
- April 1, 2013
- Report Date
- April 19, 2013
- Manufacturer
- AV-TEMECULA-CT
- Product Code
- MGB
- PMA / PMN Number
- P960043
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
(B)(4). THE DEVICE IS EXPECTED TO BE RETURNED FOR EVALUATION. IT HAS NOT YET BEEN RECEIVED. A FOLLOW UP REPORT WILL BE SUBMITTED WITH ALL ADDITIONAL RELEVANT INFORMATION. THE OTHER PROGLIDE DEVICE REFERENCED IS BEING FILED UNDER A SEPARATE MEDWATCH MFR NUMBER. IT WAS REPORTED THAT THE PROGLIDE DEVICE WAS USED IN A CALCIFIED VESSEL. PER THE INSTRUCTIONS FOR USE, THE SAFETY AND EFFECTIVENESS OF THE PROGLIDE DEVICE HAVE NOT BEEN ESTABLISHED FOR PATIENTS WITH FEMORAL ARTERY CALCIUM WHICH IS FLUOROSCOPICALLY VISIBLE AT THE ACCESS SITE.
(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED FOR EVALUATION. THE REPORTED EVENT WAS NOT CONFIRMED, BECAUSE KEY COMPONENTS WERE NOT RETURNED WITH THE DEVICE. BASED ON VISUAL AND FUNCTIONAL ANALYSIS OF THE RETURNED DEVICE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THE REPORTED EVENT. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY.
IT WAS REPORTED THAT AN ARTERIOTOMY CLOSURE OF THE MODERATELY CALCIFIED RIGHT COMMON FEMORAL ARTERY WAS ATTEMPTED USING A PROGLIDE DEVICE AFTER A PERCUTANEOUS CORONARY INTERVENTIONAL PROCEDURE. REPORTEDLY, THE SUTURE DID NOT CAPTURE THE ARTERY. A SECOND PROGLIDE DEVICE WAS USED WITH THE SAME RESULTS. THE SHEATH WAS REINSERTED AND MANUAL ARTERIAL COMPRESSION WAS APPLIED TO ACHIEVE HEMOSTASIS. THERE WAS NO REPORTED ADVERSE PATIENT SEQUELA OR CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE OR THERAPY. THE PHYSICIAN IS REPORTEDLY TRAINED IN THE USE OF THE PROGLIDE DEVICE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 204090 | PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM | SUTURE MEDIATED CLOSURE | MGB | AV-TEMECULA-CT | 30226K1 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 74 YR | Required Intervention | SHEATH: 7FANGIOMAX |