Summary report: N
LIFEPAK(R) 1000 DEFIBRILLATOR
MDR report key: 3102726
·
Received May 9, 2013
Report
- Report Number
- 3015876-2013-00376
- Event Type
- Malfunction
- Date Received
- May 9, 2013
- Date of Event
- March 28, 2013
- Report Date
- April 15, 2013
- Manufacturer
- PHYSIO-CONTROL, INC
- Product Code
- MKJ
- PMA / PMN Number
- K042404
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
(B)(4). PHYSIO-CONTROL EVALUATED THE DEVICE AND VERIFIED THE REPORTED FAILURE. THE CAUSE OF THE REPORTED FAILURE WAS OBSERVED TO BE THE SOLDER JOINTS ON LEGS 1 AND 16 OF AN INTEGRATED CIRCUIT CHIP, DESIGNATOR U1 WERE BROKEN LOOSE FROM THE ANALOG PCB ASSEMBLY. THE CUSTOMER RECEIVED A REPLACEMENT DEVICE.
Description of Event or Problem · 1
THE CUSTOMER INITIALLY REPORTED THAT THEIR DEVICE WAS SHOWING ITS SERVICE INDICATOR AND LOGGED AN EVENT CODE. AFTER AN EVALUATION OF THE DEVICE BY PHYSIO-CONTROL, IT WAS OBSERVED THAT THE DEVICE DID NOT DELIVER DEFIBRILLATION ENERGY. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED FAILURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 202927 | LIFEPAK(R) 1000 DEFIBRILLATOR | DEFIBRILLATORS, AUTOMATIC, EXTERNAL | MKJ | PHYSIO-CONTROL, INC | 1000 |