FDA Adverse Event Malfunction Summary report: N

LIFEPAK(R) 1000 DEFIBRILLATOR

MDR report key: 3102726 · Received May 9, 2013

Report

Report Number
3015876-2013-00376
Event Type
Malfunction
Date Received
May 9, 2013
Date of Event
March 28, 2013
Report Date
April 15, 2013
Manufacturer
PHYSIO-CONTROL, INC
Product Code
MKJ
PMA / PMN Number
K042404
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). PHYSIO-CONTROL EVALUATED THE DEVICE AND VERIFIED THE REPORTED FAILURE. THE CAUSE OF THE REPORTED FAILURE WAS OBSERVED TO BE THE SOLDER JOINTS ON LEGS 1 AND 16 OF AN INTEGRATED CIRCUIT CHIP, DESIGNATOR U1 WERE BROKEN LOOSE FROM THE ANALOG PCB ASSEMBLY. THE CUSTOMER RECEIVED A REPLACEMENT DEVICE.

Description of Event or Problem · 1

THE CUSTOMER INITIALLY REPORTED THAT THEIR DEVICE WAS SHOWING ITS SERVICE INDICATOR AND LOGGED AN EVENT CODE. AFTER AN EVALUATION OF THE DEVICE BY PHYSIO-CONTROL, IT WAS OBSERVED THAT THE DEVICE DID NOT DELIVER DEFIBRILLATION ENERGY. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED FAILURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
202927 LIFEPAK(R) 1000 DEFIBRILLATOR DEFIBRILLATORS, AUTOMATIC, EXTERNAL MKJ PHYSIO-CONTROL, INC 1000

Patients

Seq Age Sex Outcome Treatment
1