FDA Adverse Event Injury Summary report: N

XIA 3 TITANIUM BLOCKER

MDR report key: 3102714 · Received May 9, 2013

Report

Report Number
3005525032-2013-00040
Event Type
Injury
Date Received
May 9, 2013
Date of Event
April 9, 2013
Report Date
April 9, 2013
Manufacturer
STRYKER SPINE-SWITZERLAND
Product Code
MNH
PMA / PMN Number
D026665
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

NOT RETURNED.

Additional Manufacturer Narrative · 1

METHOD: COMPLAINT HISTORY REVIEW. RESULTS: THE EVENT INVOLVING THE SLIPPING OF THE XIA 3 TITANIUM BLOCKER WAS CONFIRMED VIA SALES REP. THE CUSTOMER DID NOT USE THE ANTI-TORQUE KEY OR THE INSERTER AS INSTRUCTED IN THE SURGICAL TECHNIQUE. NO FURTHER EVALUATION DUE TO THE ABSENCE OF THE PRODUCT. CONCLUSION: INVESTIGATION APPEARS TO BE SURGICAL PROCEDURE RELATED AND NOT PRODUCT RELATED. THE CAUSE OF THE FAILURE IS USER ERROR.

Description of Event or Problem · 1

IT WAS REPORTED THAT, "DURING XIA3 SURGERY, THE SURGEON TRIED TO USE THE UNIVERSAL TIGHTENER AND INSERT THE BLOCKER IN THE SCREW HEAD. HOWEVER, THE UNIVERSAL TIGHTENER WHICH ATTACHED BLOCKER SLIPPED. THE DURA MATER WAS HIT BY THE TIP OF THE UNIVERSAL TIGHTENER WHICH ATTACHED BLOCKER AT THAT TIME. AFTER SURGERY, THE SURGEON CHECKED THE MOTOR FUNCTION OF PATIENT. AND HE JUDGED THAT THERE WAS NO POSSIBILITY OF PARALYSIS. THE INSERTER OR ANTI-TORQUE KEY WAS NOT USED IN THIS CASE."

Description of Event or Problem · 1

IT WAS REPORTED THAT, "DURING XIA3 SURGERY, THE SURGEON TRIED TO USE THE UNIVERSAL TIGHTENER AND INSERT THE BLOCKER IN THE SCREW HEAD. HOWEVER THE UNIVERSAL TIGHTENER WHICH ATTACHED BLOCKER SLIPPED. THE DURA MATER WAS HIT BY THE TIP OF THE UNIVERSAL TIGHTENER WHICH ATTACHED BLOCKER AT THAT TIME. AFTER SURGERY, THE SURGEON CHECKED THE MOTOR FUNCTION OF PATIENT. AND HE JUDGED THAT THERE WAS NO POSSIBILITY OF PARALYSIS. THE INSERTER OR ANTI-TORQUE KEY WAS NOT USED IN THIS CASE."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
204013 XIA 3 TITANIUM BLOCKER SPINAL BLOCKER MNH STRYKER SPINE-SWITZERLAND UNK

Patients

Seq Age Sex Outcome Treatment
1 75 YR Required Intervention