FDA Adverse Event Malfunction Summary report: N

PANORAMA CENTRAL STATION

MDR report key: 3102686 · Received May 1, 2013

Report

Report Number
2221819-2013-01036
Event Type
Malfunction
Date Received
May 1, 2013
Date of Event
March 18, 2013
Report Date
April 30, 2013
Manufacturer
MINDRAY DS USA, INC
Product Code
MHX
PMA / PMN Number
031760
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
ME, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

COMPANY REP EVALUATED THE SYSTEM AND FOUND THE SYSTEM SOFTWARE WAS CORRUPTED. CORRECTIONS INCLUDED RELOADING OF THE SYSTEM'S SOFTWARE. SYSTEM WAS TESTED TO FACTORY'S SPECIFICATIONS.

Description of Event or Problem · 1

CUSTOMER REPORTED THE PANORAMA CENTRAL STATION SHUT DOWN AFTER THE FACILITY'S GENERATOR POWER FAILURE, WHICH MAY HAVE AFFECTED TELEMETRY MONITORING. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
190720 PANORAMA CENTRAL STATION CENTRAL MONITORING STATION MHX MINDRAY DS USA, INC

Patients

Seq Age Sex Outcome Treatment
1