FDA Adverse Event
Malfunction
Summary report: N
DUAL COOLER/HEATER
MDR report key: 3102620
·
Received May 1, 2013
Report
- Report Number
- 1828100-2013-00410
- Event Type
- Malfunction
- Date Received
- May 1, 2013
- Date of Event
- April 8, 2013
- Report Date
- April 8, 2013
- Manufacturer
- TERUMO CARDIOVASCULAR SYSTEMS CORP.
- Product Code
- DWC
- PMA / PMN Number
- PREAMEND
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
THE FIELD SVC REP (FSR) WAS ABLE TO DUPLICATE PROBLEM REPORTED. THE FSR WILL ORDER A 38/42 THERMOSTAT ASSEMBLY AND REPLACE.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING SET-UP OF THE DEVICE FOR A CARDIOPULMONARY BYPASS PROCEDURE, THE COOLER HEATER "38 DEGREE CELSIUS" MODE WAS NOT WORKING. WHEN THE OPERATOR SETS THE MODE TEMP TO 38 DEGREES, THE COOLER HEATER WOULD NOT STOP AT +/- DEGREE AS IT SHOULD. AS A RESULT, AN ALTERNATE DEVICE WAS EMPLOYED. THE SURGICAL PROCEDURE WAS COMPLETED SUCCESSFULLY, AND THERE WERE NO DELAYS, NO BLOOD LOSS OR NO ADVERSE CONSEQUENCES TO THE PT.