FDA Adverse Event Malfunction Summary report: N

DUAL COOLER/HEATER

MDR report key: 3102620 · Received May 1, 2013

Report

Report Number
1828100-2013-00410
Event Type
Malfunction
Date Received
May 1, 2013
Date of Event
April 8, 2013
Report Date
April 8, 2013
Manufacturer
TERUMO CARDIOVASCULAR SYSTEMS CORP.
Product Code
DWC
PMA / PMN Number
PREAMEND
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE FIELD SVC REP (FSR) WAS ABLE TO DUPLICATE PROBLEM REPORTED. THE FSR WILL ORDER A 38/42 THERMOSTAT ASSEMBLY AND REPLACE.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING SET-UP OF THE DEVICE FOR A CARDIOPULMONARY BYPASS PROCEDURE, THE COOLER HEATER "38 DEGREE CELSIUS" MODE WAS NOT WORKING. WHEN THE OPERATOR SETS THE MODE TEMP TO 38 DEGREES, THE COOLER HEATER WOULD NOT STOP AT +/- DEGREE AS IT SHOULD. AS A RESULT, AN ALTERNATE DEVICE WAS EMPLOYED. THE SURGICAL PROCEDURE WAS COMPLETED SUCCESSFULLY, AND THERE WERE NO DELAYS, NO BLOOD LOSS OR NO ADVERSE CONSEQUENCES TO THE PT.