FDA Adverse Event Malfunction Summary report: N

DUAL COOLER/HEATER

MDR report key: 3102524 · Received May 2, 2013

Report

Report Number
1828100-2013-00400
Event Type
Malfunction
Date Received
May 2, 2013
Date of Event
April 1, 2013
Report Date
April 8, 2013
Manufacturer
TERUMO CARDIOVASCULAR SYSTEMS CORP.
Product Code
DWC
PMA / PMN Number
PREAMEND
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE FIELD SERVICE REPRESENTATIVE (FSR) TROUBLESHOT THE REPORTED PROBLEM AND DISCOVERED POWER CORD CONNECTOR WAS FAULTY. THE POWER CORD WAS SUPPLIED BY ANOTHER MFR. THE USER FACILITY'S BIOMEDICAL ENGINEER WILL REPLACE POWER CORD.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING SET UP OF THE DEVICE FOR A CARDIOPULMONARY BYPASS PROCEDURE, THE COOLER HEATER UNIT KEPT TURNING OFF. AS A RESULT, AN ALTERNATE DEVICE WAS EMPLOYED. THE SURGICAL PROCEDURE WAS COMPLETED SUCCESSFULLY, AND THERE WERE NO DELAYS, NO BLOOD LOSS OR NO ADVERSE CONSEQUENCES TO THE PT.