FDA Adverse Event Malfunction Summary report: N

CONSTELLATION VISION SYSTEM

MDR report key: 3102345 · Received April 30, 2013

Report

Report Number
2028159-2013-00824
Event Type
Malfunction
Date Received
April 30, 2013
Date of Event
January 1, 2013
Report Date
April 2, 2013
Manufacturer
ALCON - IRVINE TECHNOLOGY CENTER
Product Code
HQC
PMA / PMN Number
K101285
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE COMPANY REPRESENTATIVE EXAMINED THE SYSTEM AND WAS UNABLE TO REPLICATE THE REPORTED PROBLEM. THE COMPANY REPRESENTATIVE FOUND ALL VACUUM AND PRESSURE LEVELS WITHIN PRODUCT SPECIFICATIONS. THE SYSTEM WAS THEN TESTED AND MET ALL PRODUCT SPECIFICATIONS. THERE WAS NO SAMPLE RETURNED FOR EVALUATION AND NO ADDITIONAL INFORMATION PROVIDED RELATED TO THIS EVENT. FOR THIS REASON, STEPS COULD NOT BE TAKEN TO REPLICATE OR CONFIRM THE REPORTED EVENT. A REVIEW OF COMPLAINTS FOR THE LAST (B)(4) DID NOT INDICATE ANY ADDITIONAL SIMILAR REPORTS FOR THIS SYSTEM. THE ROOT CAUSE COULD NOT BE CONCLUSIVELY DETERMINED. (B)(4).

Description of Event or Problem · 1

A CUSTOMER REPORTED THAT THE AIR/FLUID EXCHANGE WAS NOT WORKING DURING A VITRECTOMY PROCEDURE; THE AIR WAS GOING OUT BUT NOT IN. FOLLOWING A DELAY OF EIGHT MINUTES, THE UNIT WAS EXCHANGED AND THE CASE WAS ABLE TO BE COMPLETED WITHOUT HARM TO THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
187031 CONSTELLATION VISION SYSTEM UNIT, PHACOFRAGMENTATION HQC ALCON - IRVINE TECHNOLOGY CENTER LXT NA

Patients

Seq Age Sex Outcome Treatment
1