FDA Adverse Event Malfunction Summary report: N

CAREASSIST BED

MDR report key: 3102130 · Received April 23, 2013

Report

Report Number
9615739-2013-00114
Event Type
Malfunction
Date Received
April 23, 2013
Date of Event
March 27, 2013
Report Date
March 27, 2013
Manufacturer
HILL-ROM PLUVIGNER
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN FOUND THE BRAKE CASTER STEMS ARE CRACKED. THE TECHNICIAN REPLACED THE BRAKE CASTERS TO RESOLVE THE ISSUE.

Description of Event or Problem · 1

THE TECHNICIAN ALLEGED THAT THE BRAKE CASTERS ARE NOT HOLDING WHILE IN BRAKE MODE. NO PT IMPACT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
174145 CAREASSIST BED A/C POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM PLUVIGNER 1170

Patients

Seq Age Sex Outcome Treatment
1