FDA Adverse Event Malfunction Summary report: N

COYOTE? ES

MDR report key: 3102118 · Received May 9, 2013

Report

Report Number
2134265-2013-03100
Event Type
Malfunction
Date Received
May 9, 2013
Date of Event
April 11, 2013
Report Date
April 11, 2013
Manufacturer
BOSTON SCIENTIFIC - MAPLE GROVE
Product Code
LIT
PMA / PMN Number
K080982
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATED BY MFR: THE COYOTE ES CATHETER WAS RECEIVED WITH NO ORIGINAL PACKAGING OR OTHER DEVICES. THERE WAS CONTRAST IN THE BALLOON AND INFLATION LUMEN. MAGNIFIED INSPECTION REVEALED NO DAMAGE. THE DEVICE WAS INFLATED TO RATED BURST PRESSURE (RBP) WITH AN INFLATION DEVICE FILLED WITH WATER. THE DEVICE MAINTAINED RBP WITH NO INDICATION OF ANY LEAKS OR OTHER IRREGULARITIES. AFTER CONFIRMING THE DEVICE MAINTAINED RBP, THE DEVICE WAS DEFLATED BY APPLYING NEGATIVE PRESSURE WITH THE INFLATION DEVICE. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE WAS NOT CONFIRMED. (B)(4).

Additional Manufacturer Narrative · 1

AGE AT THE TIME OF EVENT : 18 YEARS OR OLDER. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING PERCUTANEOUS PERIPHERAL INTERVENTION PROCEDURE, A BALLOON RUPTURED OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE SEVERELY CALCIFIED AND MODERATELY TORTUOUS ANTERIOR TIBIAL ARTERY WITH THE DIAMETER OF 2MM. THE 1.5MM X 20MM X 142CM COYOTE ES BALLOON CATHETER WAS USED FOR DILATATION OF THE LESION. UPON THE FIRST INFLATION AT 6 ATMS, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS WAS GOOD.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING PERCUTANEOUS PERIPHERAL INTERVENTION PROCEDURE, A BALLOON RUPTURED OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE SEVERELY CALCIFIED AND MODERATELY TORTUOUS ANTERIOR TIBIAL ARTERY WITH THE DIAMETER OF 2MM. THE 1.5MM X 20MM X 142CM COYOTE ES BALLOON CATHETER WAS USED FOR DILATATION OF THE LESION. UPON THE FIRST INFLATION AT 6 ATMS, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS WAS GOOD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
202766 COYOTE? ES CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL LIT BOSTON SCIENTIFIC - MAPLE GROVE H74939134152010 14517488

Patients

Seq Age Sex Outcome Treatment
1 GUIDEWIRE:CRUISE| INTRODUCER SHEATH:ZEON6F