RETAIN-SLEEVE LONG F/MATRIX 5.5
Report
- Report Number
- 1719045-2013-01418
- Event Type
- Malfunction
- Date Received
- May 9, 2013
- Report Date
- April 15, 2013
- Manufacturer
- SYNTHES MONUMENT
- Product Code
- LXH
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SZ
- Reporter Occupation
- OTHER
Narratives
DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. THE DEVICE HISTORY RECORD REVIEW FOR THIS LOT HAS BEEN REQUESTED. SUBJECT DEVICE HAS BEEN RECEIVED AND IS CURRENTLY IN THE EVALUATION PROCESS. INVESTIGATION IS ON GOING; NO CONCLUSION COULD BE DRAWN.
THIS DEVICE USED FOR TREATMENT AND NOT DIAGNOSIS. MAGNUM MANUFACTURING CENTER MANUFACTURED THE HOLDING SLEEVE ¿ LONG FOR MATRIX, PART 03.632.036, AND LOT NUMBER 6508192. THE SUPPLIERS CERTIFICATE OF COMPLIANCE, DATED MARCH 4, 2011, INDICATES THE PARTS WERE MANUFACTURED TO PART 03.632.036, REVISION G AND MET THE REQUIRED SPECIFICATIONS. THE LOT WAS INSPECTED AND CONFORMED TO THE SYNTHES, TABULATED INCOMING FINAL INSPECTION SHEET NUMBER (B)(4), REVISION L, DATED MARCH 7, 2011. THERE WERE NO NON CONFORMITIES OR OTHER COMPLAINT RELATED ISSUES ASSOCIATED WITH THIS COMPLAINT. DUE TO AN UNKNOWN CAUSE, THE THREADED TIP BROKE. THE MATERIAL AND HARDNESS OF THE INNER SLEEVE WAS CONFIRMED TO BE WITHIN SPECIFICATION AND THE DIAMETER NEAR THE TIP PASSED SPECIFICATIONS. THE THREADS, THREADED TIP DEPTH AND LOCATION WERE UNABLE TO BE VERIFIED DUE TO THE DAMAGE ON THE TIP. THE PARTS ALL PASSED INSPECTION AT THE TIME OF MANUFACTURING.
ON AN UNKNOWN DATE, IT WAS DISCOVERED THAT A SCREW HAD THE THREAD BROKEN OFF AT THE DISTAL END. IT IS UNKNOWN WHERE THIS OCCURRED. NO ADDITIONAL INFORMATION IS AVAILABLE. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 202765 | RETAIN-SLEEVE LONG F/MATRIX 5.5 | LXH | SYNTHES MONUMENT | 6508192 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |