FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 3101920 · Received May 3, 2013

Report

Report Number
1720753-2013-05651
Event Type
Malfunction
Date Received
May 3, 2013
Date of Event
March 27, 2013
Report Date
May 3, 2013
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE MONITOR POWER SUPPLY VOLTAGE WAS ADJUSTED. THE SYSTEM WAS THEN FOUND TO BE OPERATING AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE LEFT MONITOR WAS GOING BLACK AND THE SYSTEM REQUIRED A REBOOT TO REGAIN FUNCTIONALITY. THIS RESULTED IN A LOSS OF THE LIVE IMAGE. THERE IS NO REPORT OF PATIENT INJURY ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
195291 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1