ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2013-06063
- Event Type
- Malfunction
- Date Received
- May 9, 2013
- Report Date
- April 11, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- PATIENT
Narratives
THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: THE DISPLAY LENS WAS FOUND TO BE CRACKED AROUND THE EDGES AND FOUND TO BE SCRATCHED. THE BATTERY COMPARTMENT WAS FOUND TO BE CRACKED AT THE THREADS. THE DIM PINK/FADED DISPLAY SCREEN WAS DUPLICATED. THE BAD DISPLAY SCREEN WAS REMOVED AND TESTING WAS COMPLETED WITH A NEW DISPLAY SCREEN, AND THE GOOD DISPLAY SCREEN IS SHOWING A STRONG CLEAR CONTRAST. UNRELATED TO THE COMPLAINT, EVALUATION REVEALED THAT THE KEYPAD SYMBOLS WERE WORN, WHICH HAS NO EFFECT ON INSULIN DELIVERY. (B)(6). DEVICE HISTORY RECORD REVIEW WAS CONDUCTED ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED THAT THE DISPLAY SCREEN WAS DIM PINK FADED. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON (B)(4) 2013.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 203280 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 20 YR |