FDA Adverse Event Malfunction Summary report: N

840 VENTILATOR

MDR report key: 3101761 · Received May 3, 2013

Report

Report Number
8020893-2013-00992
Event Type
Malfunction
Date Received
May 3, 2013
Date of Event
April 5, 2013
Report Date
April 5, 2013
Manufacturer
COVIDIEN, FORMERLY NELLCOR
Product Code
CBK
PMA / PMN Number
K970460
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
HI, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT AN 840 VENTILATOR WAS INOPERABLE WHILE IN USE ON A PATIENT. THE PATIENT WAS NOT HARMED OR INJURED AS A RESULT OF THE EVENT. THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) VERIFIED THE MALFUNCTION. THE CSE INSPECTED THE DEVICE AND REPLACED THE SOLENOID VALVE ASSEMBLY. THE UNIT PASSED EXTENDED SELF-TESTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
195006 840 VENTILATOR VENTILATOR CBK COVIDIEN, FORMERLY NELLCOR 840

Patients

Seq Age Sex Outcome Treatment
1