VA LOCKSCR Ø2.4 SELF-TAP L18 TAN
Report
- Report Number
- 2520274-2013-02464
- Event Type
- Malfunction
- Date Received
- May 9, 2013
- Date of Event
- May 15, 2012
- Report Date
- April 15, 2013
- Manufacturer
- SYNTHES USA
- Product Code
- HWC
- PMA / PMN Number
- K102694
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SZ
- Reporter Occupation
- OTHER
Narratives
DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. WITHOUT A LOT NUMBER, THE DEVICE HISTORY RECORDS REVIEW COULD NOT BE COMPLETED. SUBJECT DEVICE HAS BEEN RECEIVED AND IS CURRENTLY IN THE EVALUATION PROCESS. INVESTIGATION IS ON GOING; NO CONCLUSION COULD BE DRAWN.
THIS DEVICE USED FOR TREATMENT AND NOT DIAGNOSIS. THE INVESTIGATION OF THE COMPLAINED ARTICLES SHOWS THAT THE LOCKING THREAD AT THE SCREW HEADS IS PARTIALLY BADLY DEFORMED. THIS INDICATES CLEARLY MECHANICAL MISHANDLING DURING USE. IT IS POSSIBLE THAT TOO MUCH APPLIED MECHANICAL FORCE WHILE INSERTION MAY HAVE CAUSED THE DEFORMATIONS. THE PLATE THREADS ARE USED BUT OTHERWISE IN GOOD CONDITION. REVIEWING THE MANUFACTURING- AND MATERIAL DOCUMENTATION OF THE PLATE SHOWS CONFORMITY AS WELL. NO PRODUCT FAULT COULD BE DETECTED.
DEVICE REPORT FROM SYNTHES (B)(4) REPORTS AN EVENT IN (B)(6) AS FOLLOWS: DURING AN INSERTION OF VARIABLE ANGLE LOCKING SCREWS TO A PLATE ON (B)(6) 2012, THE TIP OF A SCREW TWISTED. THE SCREWS WERE INSERTED BY HAND. THE PLATE WAS FIXED WITH AN AIMING BLOCK. SIX SCREWS WERE TO BE INSERTED TO THE PLATE, BUT THE SURGEON WAS UNABLE TO FINISH LOCKING IN THE PROXIMAL ROW STYLOID HOLE. AN ADDITIONAL SCREW WAS INSERTED IN THE HOLE AND WAS RUN THROUGH. THE SURGEON REMOVED THE SCREW FROM THE FRACTURE PART OF THE PLATE. THE OPERATION WAS COMPLETED. THIS IS REPORT 1 OF 2 FOR COMPLAINT (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 203768 | VA LOCKSCR Ø2.4 SELF-TAP L18 TAN | HWC | SYNTHES USA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |