FDA Adverse Event Malfunction Summary report: N

840 VENTILATOR

MDR report key: 3101707 · Received May 3, 2013

Report

Report Number
8020893-2013-00995
Event Type
Malfunction
Date Received
May 3, 2013
Date of Event
April 8, 2013
Report Date
April 8, 2013
Manufacturer
COVIDIEN, FORMERLY NELLCOR
Product Code
CBK
PMA / PMN Number
K970460
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT AN 840 VENTILATOR WAS INOPERABLE. NO PATIENT INVOLVEMENT. THE COVIDIEN CUSTOMER SUPPORT SUPPORT ENGINEER (CSE) VERIFIED THE MALFUNCTION. THE CSE INSPECTED THE DEVICE AND REPLACED THE GUI CPU PCB AND GUI TO BD CABLE. THE UNIT PASSED EXTENDED SELF-TESTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
194040 840 VENTILATOR VENTILATOR CBK COVIDIEN, FORMERLY NELLCOR 840

Patients

Seq Age Sex Outcome Treatment
1