FDA Adverse Event Malfunction Summary report: N

840 VENTILATOR

MDR report key: 3101704 · Received May 3, 2013

Report

Report Number
8020893-2013-00975
Event Type
Malfunction
Date Received
May 3, 2013
Date of Event
March 30, 2013
Report Date
April 3, 2013
Manufacturer
COVIDIEN, FORMERLY NELLCOR
Product Code
CBK
PMA / PMN Number
K970460
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

RECEIVED INFORMATION STATING THAT DUE TO A VENTILATOR MALFUNCTION PATIENT WAS PLACED ON A SECOND VENTILATOR. THE PATIENT WAS NOT HARMED OR INJURED AS A RESULT OF THE EVENT. THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) VERIFIED THE MALFUNCTION. THE CSE INSPECTED THE DEVICE AND REPLACED THE SOLENOID VALVE ASSEMBLY. THE UNIT PASSED EXTENDED SELF TESTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
194039 840 VENTILATOR VENTILATOR CBK COVIDIEN, FORMERLY NELLCOR 840

Patients

Seq Age Sex Outcome Treatment
1