Report Number
1720753-2013-05491
Event Type
Malfunction
Date Received
April 30, 2013
Date of Event
April 9, 2013
Report Date
April 30, 2013
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE REPRESENTATIVE EVALUATED THE SYSTEM AND REPLACED A PAIR OF BATTERIES. THE SYSTEM OPERATES AS INTENDED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE SYSTEM DISPLAYED FILAMENT AND PRECHARGE ERRORS AND WOULD NOT COMPLETE BOOT UP. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
188038 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1