Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE X-RAY CONTROLLER BOARD WAS REPLACED AND THE SYSTEM SOFTWARE WAS RELOADED. THE SYSTEM WAS THEN FOUND TO BE OPERATING AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM WOULD NOT BOOT UP. THERE IS NO REPORT OF PT INJURY ASSOCIATED THIS EVENT.