FDA Adverse Event Malfunction Summary report: N

TOTAL CARE BED

MDR report key: 3101669 · Received April 30, 2013

Report

Report Number
1824206-2013-02413
Event Type
Malfunction
Date Received
April 30, 2013
Date of Event
April 2, 2013
Report Date
April 2, 2013
Manufacturer
HILL-ROM, INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN FOUND A BENT SIDE RAIL MOUNT AND D-PIN. THE TECHNICIAN REPLACED THE SIDE RAIL MOUNT AND D-PIN TO RESOLVE THE ISSUE.

Description of Event or Problem · 1

THE TECHNICIAN ALLEGED THE SIDE RAIL WOULD NOT LATCH. NO PATIENT IMPACT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
188213 TOTAL CARE BED A/C POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM, INC. 1900

Patients

Seq Age Sex Outcome Treatment
1