FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 3101603 · Received May 1, 2013

Report

Report Number
1720753-2013-05545
Event Type
Malfunction
Date Received
May 1, 2013
Date of Event
April 16, 2013
Report Date
May 1, 2013
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE HARD DRIVE, CABLE AND THE CPU UPGRADE WERE INSTALLED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THE SYSTEM WOULD LOCK UP DURING THE BOOT PROCESS. THERE IS NO REPORT OF PT INJURY OR DEATH.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
190860 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1