840 VENTILATOR

MDR report key: 3101602 · Received May 3, 2013

Report

Report Number
8020893-2013-00998
Event Type
Malfunction
Date Received
May 3, 2013
Date of Event
April 5, 2013
Report Date
April 5, 2013
Manufacturer
COVIDIEN, FORMERLY NELLCOR PURITAN
Product Code
CBK
PMA / PMN Number
K970460
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

THE SERVICE REPORT SHOWS THE CUSTOMER REPORTED THAT THE 840 VENTILATOR HAD A BLANK SCREEN. THERE WAS NO PT INVOLVEMENT. THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) INSPECTED THE DEVICE AND REPLACED THE CABLE ASSEMBLY AND THE POWER IN/OUT PCB. THE UNIT PASSED EXTENDED SELF-TESTING.