FDA Adverse Event
Malfunction
Summary report: N
9900
MDR report key: 3101597
·
Received May 1, 2013
Report
- Report Number
- 1720753-2013-05527
- Event Type
- Malfunction
- Date Received
- May 1, 2013
- Date of Event
- April 22, 2013
- Report Date
- May 1, 2013
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
A GE SVC REP PERFORMED AN ONSITE INVESTIGATION. THE GENERATOR INTERFACE BOARD WAS REPLACED AND THE CALIBRATION FILES WERE RELOADED. THE SYSTEM WAS THEN FOUND TO BE OPERATING AS INTENDED.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE SYS DISPLAYED AN X-RAY DISABLED ERROR MESSAGE THAT PREVENTED THE SYS FROM BOOTING UP. THERE IS NO REPORT OF PT INJURY ASSOCIATED WITH THE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 190681 | 9900 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9900 |