FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 3101597 · Received May 1, 2013

Report

Report Number
1720753-2013-05527
Event Type
Malfunction
Date Received
May 1, 2013
Date of Event
April 22, 2013
Report Date
May 1, 2013
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

A GE SVC REP PERFORMED AN ONSITE INVESTIGATION. THE GENERATOR INTERFACE BOARD WAS REPLACED AND THE CALIBRATION FILES WERE RELOADED. THE SYSTEM WAS THEN FOUND TO BE OPERATING AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYS DISPLAYED AN X-RAY DISABLED ERROR MESSAGE THAT PREVENTED THE SYS FROM BOOTING UP. THERE IS NO REPORT OF PT INJURY ASSOCIATED WITH THE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
190681 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1