FDA Adverse Event
Malfunction
Summary report: N
2800
MDR report key: 3101528
·
Received May 2, 2013
Report
- Report Number
- 1720753-2013-05633
- Event Type
- Malfunction
- Date Received
- May 2, 2013
- Date of Event
- April 25, 2013
- Report Date
- May 2, 2013
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE REPORTED ISSUE COULD NOT BE DUPLICATED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED: "GENERATOR ERROR. CANNOT FLUORO." THIS MESSAGE APPEARS WHEN THE SYSTEM DETECTS AN ERROR IN ANY OF THE REGISTERS ASSOCIATED WITH THE HIGH VOLTAGE GENERATOR. THE SYSTEM AUTOMATICALLY SHUTS DOWN WHEN THIS OCCURS. THERE WAS NO PATIENT INJURY OR DEATH REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 191811 | 2800 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 2800 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |