FDA Adverse Event
Malfunction
Summary report: N
MAESTRO DRILL WITH HANDSWITCH
MDR report key: 3101499
·
Received May 9, 2013
Report
- Report Number
- 0001811755-2013-01065
- Event Type
- Malfunction
- Date Received
- May 9, 2013
- Date of Event
- April 15, 2013
- Report Date
- April 15, 2013
- Manufacturer
- STRYKER INSTRUMENTS-KALAMAZOO
- Product Code
- HBB
- PMA / PMN Number
- K041754
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
ADDITIONAL INFORMATION WILL BE SUBMITTED ONCE THE DEVICE EVALUATION IS COMPLETE.
Additional Manufacturer Narrative · 1
THE COMPLAINT WAS CONFIRMED FOR RUN ON BY THE REPAIR TECHNICIAN THROUGH FUNCTIONAL TESTING, DISASSEMBLY AND VISUAL INSPECTION. THE NEEDLE VALVE WAS UNPRESSING FROM THE HOUSING. THE VALVE LIKELY BECAME DISLODGED DUE TO WEAR.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE MAESTRO DRILL WITH HANDSWITCH WAS RUNNING WITHOUT USER ACTIVATION DURING PERFORMANCE TESTING. NO ADVERSE EVENT WAS ASSOCIATED WITH THE DEVICE.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE MAESTRO DRILL WITH HANDSWITCH WAS RUNNING WITHOUT USER ACTIVATION DURING PERFORMANCE TESTING. NO ADVERSE EVENT WAS ASSOCIATED WITH THE DEVICE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 204144 | MAESTRO DRILL WITH HANDSWITCH | PNEUMATIC CRANIAL DRILL MOTOR | HBB | STRYKER INSTRUMENTS-KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |