FDA Adverse Event Malfunction Summary report: N

MAESTRO DRILL WITH HANDSWITCH

MDR report key: 3101499 · Received May 9, 2013

Report

Report Number
0001811755-2013-01065
Event Type
Malfunction
Date Received
May 9, 2013
Date of Event
April 15, 2013
Report Date
April 15, 2013
Manufacturer
STRYKER INSTRUMENTS-KALAMAZOO
Product Code
HBB
PMA / PMN Number
K041754
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ADDITIONAL INFORMATION WILL BE SUBMITTED ONCE THE DEVICE EVALUATION IS COMPLETE.

Additional Manufacturer Narrative · 1

THE COMPLAINT WAS CONFIRMED FOR RUN ON BY THE REPAIR TECHNICIAN THROUGH FUNCTIONAL TESTING, DISASSEMBLY AND VISUAL INSPECTION. THE NEEDLE VALVE WAS UNPRESSING FROM THE HOUSING. THE VALVE LIKELY BECAME DISLODGED DUE TO WEAR.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE MAESTRO DRILL WITH HANDSWITCH WAS RUNNING WITHOUT USER ACTIVATION DURING PERFORMANCE TESTING. NO ADVERSE EVENT WAS ASSOCIATED WITH THE DEVICE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE MAESTRO DRILL WITH HANDSWITCH WAS RUNNING WITHOUT USER ACTIVATION DURING PERFORMANCE TESTING. NO ADVERSE EVENT WAS ASSOCIATED WITH THE DEVICE.