Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE REP EVALUATED THE SYSTEM AND REPLACED THE POWER CAP MODULE AND THE BATTERY CHARGER CABLE. THE SYSTEM OPERATES AS INTENDED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE SYSTEM DISPLAYED A PRECHARGE ERROR AND WOULD NOT COMPLETE BOOT UP. NO PT INJURY WAS REPORTED.