FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 3101458 · Received May 2, 2013

Report

Report Number
1720753-2013-05585
Event Type
Malfunction
Date Received
May 2, 2013
Date of Event
April 25, 2013
Report Date
May 2, 2013
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. ALL MAINFRAME CIRCUIT BOARDS WERE RESEATED. THE SYSTEM WAS THEN FOUND TO BE OPERATING AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT DURING A CASE, THE DISPLAYED LIVE IMAGE BECAME DARK AND UNUSABLE. THIS RESULTED IN A LOSS OF THE LIVE IMAGE. THERE IS NO REPORT OF PATIENT INJURY ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
193162 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1