GREENLIGHT HPS BPH FIBER OPTIC
Report
- Report Number
- 2937094-2013-00537
- Event Type
- Malfunction
- Date Received
- May 3, 2013
- Date of Event
- February 12, 2013
- Report Date
- February 26, 2013
- Manufacturer
- AMERICAN MEDICAL SYSTEMS
- Product Code
- GEX
- PMA / PMN Number
- K062719
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- PHYSICIAN
Narratives
FIBER ANALYSIS: THE FIBER CAP WAS FOUND TO BE ATTACHED AND INTACT, HOWEVER DRILLED THROUGH AND EXHIBITING DETRITUS. THE FIBER CAP CONDITION WOULD PREVENT PROPER FIBER FUNCTION; LIKELY RESULTING IN A DIFFUSE BEAM AND REDUCED TISSUE VAPORIZATION EFFICIENCY. THE DRILLED THROUGH FIBER CAP CONDITION MAY ALSO RESULT IN A FORWARD FIRING CONDITION OF THE SIDE-FIRING SURGICAL FIBER. THE MOST PROBABLE ROOT CAUSE OF THE DEVICE FAILURE WAS DETERMINED TO BE HEAT ACCUMULATION/USER HANDLING DUE TO TISSUE CONTACT AND OR TECHNIQUE AND ANATOMICAL/PROCEDURAL FACTORS ENCOUNTERED DURING THE PROCEDURE, LIMITING THE PERFORMANCE OF THE FIBER.
IT WAS REPORTED THAT WHILE USING THE SURGICAL FIBER DURING A PROSTATE PROCEDURE, THE TIP WAS DAMAGED AT 178,023 JOULES OF USE. THE CASE WAS COMPLETED USING A SECOND FIBER. THERE WAS NO INJURY REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 195040 | GREENLIGHT HPS BPH FIBER OPTIC | POWERED SURGICAL LASER INSTRUMENT | GEX | AMERICAN MEDICAL SYSTEMS | 10-2090 | 242H |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |