FDA Adverse Event Malfunction Summary report: N

GREENLIGHT HPS BPH FIBER OPTIC

MDR report key: 3101424 · Received May 3, 2013

Report

Report Number
2937094-2013-00537
Event Type
Malfunction
Date Received
May 3, 2013
Date of Event
February 12, 2013
Report Date
February 26, 2013
Manufacturer
AMERICAN MEDICAL SYSTEMS
Product Code
GEX
PMA / PMN Number
K062719
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

FIBER ANALYSIS: THE FIBER CAP WAS FOUND TO BE ATTACHED AND INTACT, HOWEVER DRILLED THROUGH AND EXHIBITING DETRITUS. THE FIBER CAP CONDITION WOULD PREVENT PROPER FIBER FUNCTION; LIKELY RESULTING IN A DIFFUSE BEAM AND REDUCED TISSUE VAPORIZATION EFFICIENCY. THE DRILLED THROUGH FIBER CAP CONDITION MAY ALSO RESULT IN A FORWARD FIRING CONDITION OF THE SIDE-FIRING SURGICAL FIBER. THE MOST PROBABLE ROOT CAUSE OF THE DEVICE FAILURE WAS DETERMINED TO BE HEAT ACCUMULATION/USER HANDLING DUE TO TISSUE CONTACT AND OR TECHNIQUE AND ANATOMICAL/PROCEDURAL FACTORS ENCOUNTERED DURING THE PROCEDURE, LIMITING THE PERFORMANCE OF THE FIBER.

Description of Event or Problem · 1

IT WAS REPORTED THAT WHILE USING THE SURGICAL FIBER DURING A PROSTATE PROCEDURE, THE TIP WAS DAMAGED AT 178,023 JOULES OF USE. THE CASE WAS COMPLETED USING A SECOND FIBER. THERE WAS NO INJURY REPORTED.