FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 3101380 · Received May 2, 2013

Report

Report Number
1720753-2013-05576
Event Type
Malfunction
Date Received
May 2, 2013
Date of Event
January 17, 2013
Report Date
May 2, 2013
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ON SITE INVESTIGATION. THE HARD DRIVE WAS REPLACED AND THE SYSTEM SOFTWARE WAS RELOADED. THE SYSTEM WAS THEN FOUND TO BE OPERATING AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM DISPLAYED A FILE SYSTEM IS CORRUPTED ERROR MESSAGE AND WOULD NOT BOOT UP. THERE IS NO REPORT OF PT INJURY ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
192669 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1