FDA Adverse Event Malfunction Summary report: N

GREENLIGHT HPS BPH FIBER OPTIC

MDR report key: 3101348 · Received May 3, 2013

Report

Report Number
2937094-2013-00526
Event Type
Malfunction
Date Received
May 3, 2013
Date of Event
January 14, 2013
Report Date
February 5, 2012
Manufacturer
AMERICAN MEDICAL SYSTEMS
Product Code
GEX
PMA / PMN Number
K062719
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

FIBER ANALYSIS: THE FIBER CAP WAS FOUND TO BE ATTACHED AND INTACT, HOWEVER DRILLED THROUGH. THE CAP WAS ALSO FOUND TO EXHIBIT DETRITUS AND DEVITRIFICATION. THE FIBER CAP CONDITION WOULD PREVENT PROPER FIBER FUNCTION; LIKELY RESULTING IN A DIFFUSE BEAM AND REDUCED TISSUE VAPORIZATION EFFICIENCY. THE DRILLED THROUGH FIBER CAP CONDITION MAY ALSO RESULT IN A FORWARD FIRING CONDITION OF THE SIDE-FIRING SURGICAL FIBER. THE MOST PROBABLE ROOT CAUSE OF THE DEVICE FAILURE WAS DETERMINED TO BE HEAT ACCUMULATION DUE TO TISSUE CONTACT AND OR TECHNIQUE AND ANATOMICAL/PROCEDURAL FACTORS ENCOUNTERED DURING THE PROCEDURE, LIMITING THE PERFORMANCE OF THE FIBER.

Description of Event or Problem · 1

IT WAS REPORTED THAT WHILE USING THE SURGICAL FIBER DURING A PROSTATE PROCEDURE, THE FIBER TIP WAS DAMAGED AT 51,402 JOULES OF USE. THE CASE WAS COMPLETED USING A SECOND FIBER. THERE WAS NO INJURY REPORTED.