FDA Adverse Event
Malfunction
Summary report: N
ELLIPSE DR, DF4 CONNECTOR
MDR report key: 3101182
·
Received May 9, 2013
Report
- Report Number
- 2938836-2013-01958
- Event Type
- Malfunction
- Date Received
- May 9, 2013
- Date of Event
- February 12, 2013
- Manufacturer
- ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
- Product Code
- LWS
- PMA / PMN Number
- P910023
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.
Description of Event or Problem · 1
IT WAS REPORTED THAT AN ASYMPTOMATIC PATIENT PRESENTED TO THE CLINIC FOR A ROUTINE FOLLOW UP. IT WAS OBSERVED THAT THE DEVICE SHOWED T WAVE OVER SENSING IN THE STORED EGM. THE DEVICE WAS REPROGRAMMED AND REMAINED IMPLANTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 204595 | ELLIPSE DR, DF4 CONNECTOR | IMPLANTABLE CARDIOVERTER DEFIBRILLATOR | LWS | ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION | CD2311-36Q | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 81 YR |