FDA Adverse Event Injury Summary report: N

DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR

MDR report key: 3101085 · Received May 9, 2013

Report

Report Number
2938836-2013-01877
Event Type
Injury
Date Received
May 9, 2013
Date of Event
February 7, 2013
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. DEVICE EVALUATION ANTICIPATED BUT NOT YET BEGUN.

Additional Manufacturer Narrative · 1

THE DAMAGE FOUND WAS SUSTAINED DURING THE SURGICAL PROCEDURE. THE LEAD WAS OTHERWISE NORMAL.

Description of Event or Problem · 1

IT WAS REPORTED THAT AN INCREASE IN THRESHOLD WAS NOTED. REVIEW OF DIAGNOSTIC IMAGE SHOWED THE LEAD HAS EXTENDED AND STRETCHED. AFTER ATTEMPTED UNSUCCESSFUL REPOSITIONING, THE LEAD WAS EXPLANTED AND REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
203852 DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR DEFIBRILLATION LEAD NVY ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 7120Q/58 NA

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention