FDA Adverse Event Malfunction Summary report: N

VENTED AUTOFEED CHAMBER

MDR report key: 3100868 · Received May 8, 2013

Report

Report Number
9611451-2013-00362
Event Type
Malfunction
Date Received
May 8, 2013
Date of Event
April 9, 2013
Report Date
April 11, 2013
Manufacturer
FISHER & PAYKEL HEALTHCARE LTD
Product Code
BTT
PMA / PMN Number
K934140
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

(B)(4). METHOD: THE COMPLAINT MR290V CHAMBER WAS RETURNED TO FISHER & PAYKEL HEALTHCARE IN (B)(4) AND WAS VISUALLY INSPECTED. RESULTS: VISUAL INSPECTION REVEALED THAT THE RETURNED DEVICE HAD A BREAK IN THE FEEDSET TUBING AT THE CONNECTION TO THE CHAMBER. THE SURFACE OF THE BREAK WAS ROUGH AND NOT SMOOTHLY CUT. A LOT CHECK REVEALED NO OTHER COMPLAINT OF THIS TYPE FOR LOT NUMBER 120821. CONCLUSION: THE DAMAGE APPEARED TO BE CAUSED BY THE FEEDSET TUBE BEING PULLED AWAY FROM THE CHAMBER, POSSIBLY DUE TO THE FEEDSET TUBE BEING CAUGHT OR UNDER TENSION. WE HAVE CONDUCTED EXTENSIVE TESTING OF OUR MR290 CHAMBER, WITH PARTICULAR EMPHASIS ON FEEDSET BREAKS. (B)(4). IN ADDITION, ALL CHAMBERS ARE PRESSURE TESTED BEFORE THEY LEAVE THE PRODUCTION LINE AND ANY HOLES OR LEAKS IN THE FEEDSET ARE IDENTIFIED DURING THIS PROCESS. THIS SUGGESTS THAT THE DAMAGE TO THE FEEDLINE TUBE OCCURRED AFTER THE PRODUCT WAS RELEASED FOR DISTRIBUTION. THE USER INSTRUCTIONS WHICH ACCOMPANY THE MR290 CHAMBER STATE THE FOLLOWING: "SET APPROPRIATE VENTILATOR ALARMS." "PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT." (B)(4).

Additional Manufacturer Narrative · 1

(B)(4). THE COMPLAINT MR290 DEVICE IS CURRENTLY IN TRANSIT TO FISHER & PAYKEL HEALTHCARE (B)(4) FOR EVALUATION. WE WILL PROVIDE A FOLLOW UP REPORT UPON RECEIPT OF THE COMPLAINT DEVICE AND COMPLETION OF OUR INVESTIGATION.

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED VIA A DISTRIBUTOR THAT A WATER LEAK OCCURRED AT THE FEEDSET OF AN MR290 AUTOFEED HUMIDIFICATION CHAMBER. THIS WAS FOUND PRIOR TO PATIENT USE.

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED VIA A DISTRIBUTOR THAT A WATER LEAK OCCURRED AT THE FEEDSET OF AN MR290 AUTOFEED HUMIDIFICATION CHAMBER. THIS WAS FOUND PRIOR TO PATIENT USE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
201407 VENTED AUTOFEED CHAMBER AUTOFEED CHAMBER BTT FISHER & PAYKEL HEALTHCARE LTD MR290V 1208210105

Patients

Seq Age Sex Outcome Treatment
1