ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2013-06054
- Event Type
- Malfunction
- Date Received
- May 8, 2013
- Report Date
- April 13, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- OTHER
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
FOLLOW-UP #1 (B)(4) ¿ DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: THERE WAS NO DATA FROM THE DATE OF THE EVENT DUE TO CONTINUED PATIENT USE. THE HISTORY SHOWED THAT THE TIME AND DATE WERE MANUALLY MANIPULATED ON (B)(4) 2013 LEADING THE TOTAL DAILY DOSE TO RECORD THAT DATE TWICE. THE PUMP WAS EXERCISED FOR 5 DAYS AND CONFIRMED THAT NO TIME CHANGES OCCURRED DURING TESTING. THE PUMP WAS POWERED OFF FOR 3 HOURS AND THE PUMP TIME AND DATE WERE CONFIRMED TO BE MAINTAINED ACCURATELY IN THE PUMP. NO PUMP DEFECTS WERE FOUND RELATED TO THE EVENT. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
THE REPORTER CONTACTED ANIMAS ON (B)(6) 2013 AND WHILE TROUBLESHOOTING AN UNRELATED ISSUE NOTED THAT THE TIME AND DATE WERE SET INCORRECTLY IN THE PUMP. THE TIME/DATE AT THE TIME OF THE CALL WAS 5:40PM (B)(6) AND PUMP WAS SET AT 5:40AM (B)(6). THERE IS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE TIME/DATE ISSUE WASN'T RESOLVED WITH TROUBLESHOOTING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 200723 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 47 YR |