MEGA SUTURECUT NEEDLE DRIVER INSTRUMENT
Report
- Report Number
- 2955842-2013-01567
- Event Type
- Malfunction
- Date Received
- May 8, 2013
- Date of Event
- April 11, 2013
- Report Date
- April 12, 2013
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K050369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL
- Reporter Occupation
- OTHER
Narratives
THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING EVALUATION FOUND THAT THE INSTRUMENT HAD A BROKEN GRIP CABLE. THE ONE GRIP CLOSE CABLE WAS BROKEN AT THE DISTAL IDLERS. THE IDLER PULLEY SPUN FREELY AND WAS UNDAMAGED. THE CABLE SEGMENT STUCK OUT AT THE WRIST. THE OTHER CABLES AT WRIST WERE UNDAMAGED. ENGINEERING ALSO FOUND SCRATCHES ON THE SURFACE OF THE DISTAL PULLEY. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER, THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.
IT WAS REPORTED THAT DURING A DA VINCI SI CHOLECYSTECTOMY PROCEDURE THE MEGA SUTURECUT NEEDLE DRIVER INSTRUMENT WAS NOTED TO HAVE A FRAYED CABLE ON THE DISTAL END. NO MISSING OR FALLEN PIECES WERE REPORTED. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 201524 | MEGA SUTURECUT NEEDLE DRIVER INSTRUMENT | ENDOSCOPIC INSTRUMENT | NAY | INTUITIVE SURGICAL,INC. | 420309-01 | M10120608 088 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | DA VINCI SI SYSTEM, INSTRUMENTS AND ACCESSORIES |