PARADIGM REAL-TIME INSULIN INFUSION PUMP
Report
- Report Number
- 3004209178-2013-93265
- Event Type
- Malfunction
- Date Received
- May 8, 2013
- Date of Event
- May 5, 2013
- Report Date
- May 5, 2013
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
NO BUTTON ERROR ALARMS NOTED. THE INSULIN PUMP HAD NO BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES. NO NUMBER RAMPING OR SCROLLING SCREENS NOTED. USED A TEST KEYPAD, THE INSULIN PUMP RESPONDED PROPERLY TO BUTTON PRESSES AND THE TIME ADVANCED. NO FROZEN SCREEN NOTED. UNABLE TO PERFORM THE DISPLACEMENT TEST, REWIND, BASIC OCCLUSION TEST, OCCLUSION TEST, PRIME TEST AND THE EXCESSIVE NO DELIVERY TEST DUE TO BUTTON/KEYPAD ANOMALY. THE INSULIN PUMP HAD A SCRATCHED DISPLAY WINDOW.
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE PARADIGM REAL-TIME VEO INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES. IT WAS REPORTED THAT THE INSULIN PUMP HAD A FROZEN DISPLAY. CALLER STATED THAT THE TIME ON THE INSULIN PUMP IS NOT ADVANCING AND THERE IS NO RESPONSE FROM ANY BUTTON. NOTHING FURTHER WAS REPORTED.
IT WAS REPORTED THAT THE INSULIN PUMP HAD A FROZEN DISPLAY. CALLER STATED THAT THE TIME ON THE INSULIN PUMP IS NOT ADVANCING AND THERE IS NO RESPONSE FROM ANY BUTTON. NOTHING FURTHER WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 201746 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-754CMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |