FDA Adverse Event Malfunction Summary report: N

MEGA SUTURECUT NEEDLE DRIVER INSTRUMENT

MDR report key: 3100766 · Received May 8, 2013

Report

Report Number
2955842-2013-01566
Event Type
Malfunction
Date Received
May 8, 2013
Date of Event
April 9, 2013
Report Date
April 11, 2013
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. ENGINEERING WAS UNABLE TO CONFIRM THE CUSTOMER REPORTED COMPLAINT. ENGINEERING PERFORMED VISUAL INSPECTION, AND FOUND THE TIPS WERE PERFECTLY ALIGNED. ENGINEERING ALSO FOUND A BROKEN GRIP CABLE. THE ONE GRIP CLOSE CABLE WAS BROKEN AT THE DISTAL IDLERS. THE IDLER PULLEY SPUN FREELY AND WAS UNDAMAGED. THE CABLE SEGMENT STUCK OUT AT THE WRIST. THE OTHER CABLES AT THE WRIST WERE UNDAMAGED. NO OTHER DAMAGE WAS FOUND. ENGINEERING ALSO FOUND AN INDENTATION ON THE EDGE OF THE DISTAL PULLEY. ENGINEERING CONCLUDED THE DAMAGE WAS LIKELY DUE TO MISHANDLING. THE DISTAL END OF THE MAIN TUBE HAD VARIOUS SCRATCH MARKS SHOWING LIGHT MATERIAL REMOVAL. THE SCRATCHES WERE SHORT IN LENGTH AND NOT AXIALLY ALIGNED WITH THE TUBE. ENGINEERING CONCLUDED THAT DAMAGE WAS LIKELY DUE TO MISHANDLING. THE INSTRUMENTS AND ACCESSORIES USER MANUAL SPECIFICALLY STATES: GENERAL PRECAUTIONS AND WARNINGS O HANDLE INSTRUMENTS WITH CARE. AVOID MECHANICAL SHOCK OR STRESS THAT CAN CAUSE DAMAGE TO THE INSTRUMENTS. DO NOT USE AN INSTRUMENT TO CLEAN DEBRIS FROM ANOTHER INSTRUMENT INTRAOPERATIVELY. THIS MAY RESULT IN DAMAGE TO THE INSTRUMENTS OR OTHER UNINTENDED CONSEQUENCES, SUCH AS DISCONNECTION OF THE INSTRUMENT TIP. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER, THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT PRIOR TO STARTING A DA VINCI SI SURGICAL PROCEDURE THE MEGA SUTURECUT NEEDLE DRIVER INSTRUMENT WAS NOTED TO HAVE A BENT TIP. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
200536 MEGA SUTURECUT NEEDLE DRIVER INSTRUMENT ENDOSCOPIC INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420309-01 M10130207 208

Patients

Seq Age Sex Outcome Treatment
1 DA VINCI SI SYSTEM, INSTRUMENTS AND ACCESSORIES