FDA Adverse Event
Malfunction
Summary report: N
RESERVOIR 3ML
MDR report key: 3100728
·
Received May 8, 2013
Report
- Report Number
- 3004209178-2013-93251
- Event Type
- Malfunction
- Date Received
- May 8, 2013
- Date of Event
- May 4, 2013
- Report Date
- May 4, 2013
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- FRN
- PMA / PMN Number
- K032005
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE THEREFORE CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE.
Description of Event or Problem · 1
IT IS REPORTED THAT THE RESERVOIR LEAKED. CUSTOMER'S BLOOD GLUCOSE READING IS 13.5 MMOL/L. CUSTOMER IS IN THE EMERGENCY ROOM. CUSTOMER STATED THAT THE RESERVOIR LEAKED PAST BOTH O-RINGS. NOTHING FURTHER REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 200427 | RESERVOIR 3ML | RESERVOIR | FRN | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-332A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |