FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
MDR report key: 3100707
·
Received May 8, 2013
Report
- Report Number
- 3004209178-2013-93230
- Event Type
- Malfunction
- Date Received
- May 8, 2013
- Date of Event
- April 11, 2013
- Report Date
- April 11, 2013
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
THE INSULIN PUMP WAS RECEIVED WITH PRIME ALARM DURING BASIC OCCLUSION TEST AND UNABLE TO PRIME DURING PRIME TEST DUE TO PROTRUDED/LOOSE DRIVE SUPPORT DISK.
Description of Event or Problem · 1
CUSTOMER STATED THAT THE INSULIN PUMP IS STUCK IN THE PRIME LOOP AND INSULIN IS SQUIRTING OUT. CUSTOMER'S BLOOD GLUCOSE READING IS 123 MG/DL. CUSTOMER STATED THAT THE DRIVE SUPPORT DISK IS PROTRUDING. ADVISED CUSTOMER THAT THE INSULIN PUMP WILL NEED TO BE REPLACED. NOTHING FURTHER REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 200404 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-723NAB |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 18 YR |