FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 3100707 · Received May 8, 2013

Report

Report Number
3004209178-2013-93230
Event Type
Malfunction
Date Received
May 8, 2013
Date of Event
April 11, 2013
Report Date
April 11, 2013
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP WAS RECEIVED WITH PRIME ALARM DURING BASIC OCCLUSION TEST AND UNABLE TO PRIME DURING PRIME TEST DUE TO PROTRUDED/LOOSE DRIVE SUPPORT DISK.

Description of Event or Problem · 1

CUSTOMER STATED THAT THE INSULIN PUMP IS STUCK IN THE PRIME LOOP AND INSULIN IS SQUIRTING OUT. CUSTOMER'S BLOOD GLUCOSE READING IS 123 MG/DL. CUSTOMER STATED THAT THE DRIVE SUPPORT DISK IS PROTRUDING. ADVISED CUSTOMER THAT THE INSULIN PUMP WILL NEED TO BE REPLACED. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
200404 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723NAB

Patients

Seq Age Sex Outcome Treatment
1 18 YR