FDA Adverse Event Injury Summary report: N

VITALITY 2

MDR report key: 3100193 · Received May 8, 2013

Report

Report Number
2124215-2013-07639
Event Type
Injury
Date Received
May 8, 2013
Date of Event
September 11, 2012
Report Date
May 8, 2013
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
LWS
PMA / PMN Number
P960040
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE PRODUCT HAS BEEN RECEIVED FOR ANALYSIS. THIS REPORT WILL BE UPDATED UPON COMPLETION OF ANALYSIS.

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE DEVICE WAS PERFORMED. A VISUAL INSPECTION OF THE DEVICE HEADER AND CASE NOTED NO ANOMALIES. THE DEVICE WAS THEN EXPOSED TO SIMULATED HEART LOAD CONDITIONS, AND THE DEFIBRILLATION, PACING AND SENSING FUNCTIONS WERE TESTED. THE DEVICE OPERATED APPROPRIATELY, ACCORDING TO ITS PERFORMANCE SPECIFICATIONS, WITH NO INTERRUPTIONS IN THERAPY OUTPUT OR PROGRAMMER COMMUNICATION AT THE RETURNED PROGRAMMED SETTINGS. A SERIES OF ELECTRICAL TESTS WAS ALSO PERFORMED, AND AGAIN, NORMAL DEVICE FUNCTION WAS OBSERVED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) AND IMPLANTABLE DEFIBRILLATION LEAD EXHIBITED NOISE ON THE RATE/SENSE CHANNEL. THE NOISE WAS OVERSENSED RESULTING IN NONSUSTAINED VENTRICULAR TACHYCARDIA (VT) EPISODES. PACING WAS NOT INHIBITED AS THE PATIENT HAS A NORMAL SINUS RHYTHM. HIGH OUT OF RANGE SHOCK IMPEDANCE MEASUREMENTS WERE LATER NOTED. AN INVASIVE PROCEDURE WAS PERFORMED TO REVISE THE LEAD. THE DEVICE AND LEAD WERE SUCCESSFULLY EXPLANTED AND REPLACED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
200936 VITALITY 2 IMPLANTABLE CARDIOVERTER DEFIBRILLATOR LWS GUIDANT CRM CLONMEL IRELAND T167

Patients

Seq Age Sex Outcome Treatment
1 65 YR Hospitalization| L| R 4470| 0148| T167| 624