FDA Adverse Event
Malfunction
Summary report: N
AFFINITY 4 BIRTHING BED
MDR report key: 3100176
·
Received April 29, 2013
Report
- Report Number
- 1824206-2013-02401
- Event Type
- Malfunction
- Date Received
- April 29, 2013
- Date of Event
- April 2, 2013
- Report Date
- April 2, 2013
- Manufacturer
- HILL-ROM, INC.
- Product Code
- HDD
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE ACCOUNT FOUND THE E RING CAME OFF THE D PIN THAT SECURES THE SIDE RAIL AND THIS CAUSED THE D-PIN TO FALL OUT OF THE SIDE RAIL. THE ACCOUNT REPLACED THE E RING AND D PIN ON THE SIDE RAIL TO RESOLVE THE ISSUE.
Description of Event or Problem · 1
THE ACCOUNT ALLEGED THE SIDE RAIL WOULD NOT LATCH. NO INJURY REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 184203 | AFFINITY 4 BIRTHING BED | BIRTHING BED | HDD | HILL-ROM, INC. | 3700 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|