FDA Adverse Event Malfunction Summary report: N

AFFINITY 4 BIRTHING BED

MDR report key: 3100176 · Received April 29, 2013

Report

Report Number
1824206-2013-02401
Event Type
Malfunction
Date Received
April 29, 2013
Date of Event
April 2, 2013
Report Date
April 2, 2013
Manufacturer
HILL-ROM, INC.
Product Code
HDD
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE ACCOUNT FOUND THE E RING CAME OFF THE D PIN THAT SECURES THE SIDE RAIL AND THIS CAUSED THE D-PIN TO FALL OUT OF THE SIDE RAIL. THE ACCOUNT REPLACED THE E RING AND D PIN ON THE SIDE RAIL TO RESOLVE THE ISSUE.

Description of Event or Problem · 1

THE ACCOUNT ALLEGED THE SIDE RAIL WOULD NOT LATCH. NO INJURY REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
184203 AFFINITY 4 BIRTHING BED BIRTHING BED HDD HILL-ROM, INC. 3700

Patients

Seq Age Sex Outcome Treatment
1