ANIMAS VIBE
Report
- Report Number
- 2531779-2013-06038
- Event Type
- Malfunction
- Date Received
- May 8, 2013
- Date of Event
- April 4, 2013
- Report Date
- April 12, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
FOLLOW-UP #1 DATE OF SUBMISSION (B)(4) 2013 DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: THE KEYPAD WAS FOUND TO INTACT. EVALUATION REVEALED THAT THE KEYPAD BUTTONS WERE RESPONSIVE TO USER INPUT. THERE WAS NO EVIDENCE OF CONTAMINATION FOUND UNDER THE KEY CONTACTS. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
(B)(4). THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
THE DISTRIBUTOR CONTACTED ANIMAS ON (B)(6) 2013 ALLEGING THE UP AND DOWN ARROW KEYPAD BUTTONS WERE UNRESPONSIVE. NO FURTHER INFORMATION WAS AVAILABLE. THERE WAS NO REPORTED ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE ALLEGED KEYPAD BUTTON MALFUNCTION REMAINED UNRESOLVED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 201815 | ANIMAS VIBE | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |