FDA Adverse Event Other Summary report: N

CONSEPT 1-STEP DISINFECTING SOLUTION, DEV. 1

MDR report key: 3099558 · Received April 29, 2013

Report

Report Number
3004178847-2013-00005
Event Type
Other
Date Received
April 29, 2013
Date of Event
January 1, 2013
Report Date
January 29, 2013
Manufacturer
ABBOTT MEDICAL OPTICS INC.
Product Code
LPN
PMA / PMN Number
P850088
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CHEMISTRY EVALUATIONS OF THE ACTUAL PRODUCT USED BY THE CONSUMER WERE PERFORMED; ALL RESULTS WERE WITHIN SPECIFICATIONS. PRELIMINARY TESTING PERFORMED AT (B)(4) QUALITY CONTROL LAB. ADDITIONAL PRODUCT ANALYSIS (MFG RECORD REVIEW, RETURNED-CHEMISTRY TESTING, AND RETAIN-MICROBIOLOGICAL TESTING) WILL BE PERFORMED AT MFG SITE(S). CHEMISTRY TESTING OF A RETAINED UNIT OF THE REPORTED LOT WAS COMPLETED. THE SAMPLE MET THE FOLLOWING SPECIFICATIONS: APPEARANCE, PH, TABLETS NEUTRALIZATION AND DISSOLUTION OF 9081X AND HYDROGEN PEROXIDE ASSAY. ALL AVAILABLE PERTINENT INFO HAS BEEN SUBMITTED.

Description of Event or Problem · 1

OUR OFFICE IN (B)(6) RECEIVED A REPORT (B)(6) 2013, FROM A WHOLESALE CUSTOMER. THEY REPORTED THAT A FEMALE PT IN HER TWENTIES DISINFECTED / NEUTRALIZED HER LENSES WITH CONSEPT 1-STEP, LEFT THEM IN THE NEUTRALIZED SOLUTION OVER 7 DAYS, THEN WORE LENSES. AFTER WHICH, THE PT EXPERIENCED A CORNEAL ULCER (EYE UNK). F/U WITH THE PT ON (B)(6) 2013, INDICATED THAT WHILE SHE INITIALLY STATED SHE SOAKED HER BIOFINITY CONTACT LENSES IN SOLUTION FOR OVER 7 DAYS AFTER DISINFECTION, SHE REVISED THAT TIME PERIOD TO 3 OF 4 DAYS. THE PT ALSO REPORTED THAT HER EYE CONDITION SEEMED TO BE RECOVERED A LITTLE, BUT SHE STILL HAD NOT VISITED AN EYE CLINIC. THE PT REPORTED VISITING AN EYE CLINIC (DATE AND TREATMENT UNK) AND WAS DIAGNOSED WITH A CORNEAL ULCER. THE PT DECIDED TO CHANGE THE TYPE OF LENS SHE WEARS TO A DAILY DISPOSABLE LENS. SEE ASSOCIATED REPORT, DEVICE 2: 2020664-2013-10002.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
184682 CONSEPT 1-STEP DISINFECTING SOLUTION, DEV. 1 LPN ABBOTT MEDICAL OPTICS INC. ZK00584

Patients

Seq Age Sex Outcome Treatment
1 Other BIOFINITY CONTACT LENSES| CONSEPT 1-STEP NEUTRALIZING TABLETS