Description of Event or Problem · 1

ACCOUNT REPORTS INCREASED INFECTION RATES OF CENTRAL LINES WHICH THEY ATTRIBUTE MAY BE FROM THE PREFILLED HEPARIN AND SALINE SYRINGES. CULTURES OF SOLUTIONS OF PREFILLED SYRINGES WERE NEGATIVE BUT CULTURES FROM THE OUTSIDE OF THE SYRINGES HAVE CULTURED CHYSEOMONAS LUTEOLA (35/40) AND ACINETOBACTER IWOFFII (5/40) IN THEIR LAB WHICH ARE THE SAME ORGANISMS CULTURED FROM THE CATHETERS OF APPROXIMATELY 50 PEDIATRIC HOME CARE AND INPATIENT ONCOLOGY PATIENTS. ACCOUNT DID NOT SUPPLY SPECIFIC INFORMATION, TYPE OF INFECTION OR TREATMENT ON EACH PATIENT. PRODUCTS IDENTIFIED: PRODUCT CODE/LOT NUMBER/MANUFACT. DATE 2K0905/KH00205/08/02/00, 2K0905/KH00107/02/22/00, 2K6050/B00062/05/23/00, 2K6030/B00100/07/19/00.

Description of Event or Problem · 1

ACCOUNT REPORTS INCREASED INFECTION RATES OF CENTRAL LINES WHICH THEY ATTRIBUTE MAY BE FROM THE PREFILLED HEPARIN AND SALINE SYRINGES. CULTURES OF SOLUTIONS OF PREFILLED SYRINGES WERE NEGATIVE BUT CULTURES FROM THE OUTSIDE OF THE SYRINGES HAVE CULTURED CHYSEOMONAS LUTEOLA (35/40) AND ACINETOBACTER IWOFFII (5/40) IN THEIR LAB WHICH ARE THE SAME ORGANISMS CULTURED FROM THE CATHETERS OF APPROXIMATELY 50 PEDIATRIC HOME CARE AND INPATIENT ONCOLOGY PATIENTS. ACCOUNT DID NOT SUPPLY SPECIFIC INFORMATION, TYPE OF INFECTION OR TREATMENT ON EACH PATIENT. PRODUCTS IDENTIFIED: PRODUCT CODE/LOT NUMBER/MANUFACT. DATE 2K0905/KH00205/08/02/00, 2K0905/KH00107/02/22/00, 2K6050/B00062/05/23/00, 2K6030/B00100/07/19/00.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 NORMAL SALINE PREFILLED FLUSH COMPASS PREFILLED SYRINGE FOZ BAXTER HEALTHCARE COMPASS NA NA01

Patients

Seq Age Sex Outcome Treatment
1 UNKNOWN Required Intervention