Manufacturer
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
Product Code
LWS
PMA / PMN Number
P980016
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: (B)(4) THE DEVICE WAS RETURNED AND ANALYZED. ANALYSIS REVEALED THE RETURNED DEVICE FOUND A HIGH CURRENT DRAIN CONDITION DUE TO CURRENT LEAKAGE IN A CERAMIC CAPACITOR.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE MAY HAVE HAD A PREMATURE BATTERY DEPLETION. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
197190 VIRTUOSO VR DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MEDTRONIC MED REL MEDTRONIC PUERTO RICO D154VWC

Patients

Seq Age Sex Outcome Treatment
1 00014 YR Hospitalization| R