V.A.C. THERAPY
Report
- Report Number
- 3009897021-2013-00038
- Event Type
- Injury
- Date Received
- April 30, 2013
- Date of Event
- April 5, 2013
- Report Date
- April 5, 2013
- Manufacturer
- KCI USA INC.
- Product Code
- OMP
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MD, US
- Reporter Occupation
- NURSE
Narratives
BASED ON INFO PROVIDED, IT CANNOT BE DETERMINED WHEN THE FOREIGN BODIES ALLEGED TO BE GRANUFOAM DRESSING WERE INADVERTENTLY LEFT IN THE WOUND. THE FOREIGN MATERIAL WAS NOT RETURNED TO KCI FOR IDENTIFICATION; THEREFORE, KCI IS UNABLE TO CONFIRM ITS IDENTITY. THERE HAVE BEEN SEVERAL ATTEMPTS MADE TO GATHER ADD'L INFO, BUT THERE HAS BEEN NO RESPONSE. THIS REPORT IS BEING FILED DUE TO POSSIBLE USER ERROR. DEVICE LABELING, AVAILABLE IN PRINT AND ONLINE, STATES: V.A.C. FOAM DRESSINGS ARE NOT BIOABSORBABLE. ALWAYS COUNT THE TOTAL NUMBER OF PIECES OF FOAM REMOVED FROM THE WOUND AND ENSURE THE SAME NUMBER OF PIECES OF FOAM WAS REMOVED AS PLACED. FOAM LEFT IN THE WOUND FOR GREATER THAN THE RECOMMENDED TIME PERIOD MAY FOSTER IN-GROWTH OF TISSUE INTO THE FOAM AND CREATE DIFFICULTY IN REMOVING FOAM FROM THE WOUND OR LEAD TO INFECTION OR OTHER ADVERSE EVENTS.
ON (B)(6) 2013, THE HOME HEALTH NURSE REPORTED OBSERVING SMALL PIECES OF A FOREIGN MATERIAL ALLEGED TO BE V.A.C. GRANUFOAM DRESSING ADHERING TO THE WOUND. THE PATIENT INFORMED THE NURSE THAT THE PHYSICIAN WAS PLANNING TO "DEBRIDE THEM OUT," BUT HAS YET TO REMOVE THE FOREIGN MATERIAL. ON (B)(6) 2013, THE FOLLOWING INFO WAS REPORTED TO KCI BY THE WOUND CARE NURSE: THE WOUND HAS NOT BEEN DEBRIDED AT THIS TIME. THE PATIENT NEEDS TO SCHEDULE AN APPOINTMENT IN ORDER FOR THE PHYSICIAN TO PERFORM THE DEBRIDEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 186398 | V.A.C. THERAPY | OMP | KCI USA INC. | GRANUFOAM DRESSING |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 24 YR | Required Intervention |